  • Laser wavelength

    Starter: 785 nm
    Base Model: 532 nm, 785 nm, 1000 nm
    Hybrid: 785 nm

  • Spectral coverage

    Starter:
    300-3300 cm-1 (785 nm)
    Base Model:
    150-4350 cm-1 (532 nm)
    150-3425 cm-1 (785 nm)
    200-2400 cm-1 (1000 nm)
    Hybrid: 175-1890 cm-1 (785 nm)

  • Laser wavelength

    Starter: 785 nm
    Base Model: 532 nm, 785 nm, 1000 nm
    Hybrid: 785 nm

  • Spectral coverage

    Starter:
    300-3300 cm-1 (785 nm)
    Base Model:
    150-4350 cm-1 (532 nm)
    150-3425 cm-1 (785 nm)
    200-2400 cm-1 (1000 nm)
    Hybrid: 175-1890 cm-1 (785 nm)

    • 测量原理

      Raman spectroscopy

    • Laser wavelength

      Starter: 785 nm
      Base Model: 532 nm, 785 nm, 1000 nm
      Hybrid: 785 nm

    • Spectral coverage

      Starter:
      300-3300 cm-1 (785 nm)
      Base Model:
      150-4350 cm-1 (532 nm)
      150-3425 cm-1 (785 nm)
      200-2400 cm-1 (1000 nm)
      Hybrid: 175-1890 cm-1 (785 nm)

    • Spectral resolution

      Starter:
      10 cm-1 (average)
      Base model (average):
      5 cm-1 (532 nm)
      4 cm-1 (785 nm)
      5 cm-1 (1000 nm)
      Hybrid:
      4 cm-1 (785 nm) average

    • Channels

      Starter:
      Single channel
      Base Model:
      Up to four channels
      Hybrid:
      Up to two channels

    • Temperature

      Operating:
      15 to 30 °C
      Storage:
      -15 to 50 °C

    • Relative humidity

      20-80% RH, non-condensing

    • Input voltage

      100-240 V, 50-60 Hz, ±10%

    • Power consumption (W)

      400 (max)
      250 (typical start-up)
      120 (typical running)

    • Warm up time (minutes)

      120

    • Unit dimensions (width x height x depth in mm)

      Benchtop model: 279 x 483 x 592
      Cart model: 685 x 1022 (to tabletop) x 753

    • Weight (kg)

      Base model: 32
      Cart model: 93

    • Sampling probe compatibility

      Starter:
      Raman Rxn-10 (with accessory optics)
      Base model:
      Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn 46
      Hybrid:
      Channel 1 - Raman Rxn-20 (with accessory optics)
      Channel 2 - Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn 46

    • Automation interface

      OPC
      Modbus
      HTTPS
      (contact us for other options)

    • Installation options

      Benchtop (standard) or wheeled cart

    • Hazardous area certifications

      ATEX, CSA, IECEx

    • 测量原理

      Raman spectroscopy

    • Laser wavelength

      Starter: 785 nm
      Base Model: 532 nm, 785 nm, 1000 nm
      Hybrid: 785 nm

    • Spectral coverage

      Starter:
      300-3300 cm-1 (785 nm)
      Base Model:
      150-4350 cm-1 (532 nm)
      150-3425 cm-1 (785 nm)
      200-2400 cm-1 (1000 nm)
      Hybrid: 175-1890 cm-1 (785 nm)

    • Spectral resolution

      Starter:
      10 cm-1 (average)
      Base model (average):
      5 cm-1 (532 nm)
      4 cm-1 (785 nm)
      5 cm-1 (1000 nm)
      Hybrid:
      4 cm-1 (785 nm) average

    • Channels

      Starter:
      Single channel
      Base Model:
      Up to four channels
      Hybrid:
      Up to two channels

    • Temperature

      Operating:
      15 to 30 °C
      Storage:
      -15 to 50 °C

    • Relative humidity

      20-80% RH, non-condensing

    • Input voltage

      100-240 V, 50-60 Hz, ±10%

    • Power consumption (W)

      400 (max)
      250 (typical start-up)
      120 (typical running)

    • Warm up time (minutes)

      120

    • Unit dimensions (width x height x depth in mm)

      Benchtop model: 279 x 483 x 592
      Cart model: 685 x 1022 (to tabletop) x 753

    • Weight (kg)

      Base model: 32
      Cart model: 93

    • Sampling probe compatibility

      Starter:
      Raman Rxn-10 (with accessory optics)
      Base model:
      Raman Rxn-10 (with accessory optics)
      Hybrid:
      Channel 1 - Raman Rxn-20 (with accessory optics)
      Channel 2 - Raman Rxn-10 (with accessory optics)

    • Automation interface

      OPC
      Modbus
      HTTPS
      (contact us for other options)

    • Installation options

      Benchtop (standard) or wheeled cart

    • Hazardous area certifications

      ATEX, CSA, IECEx

    • 测量原理

      Raman spectroscopy

    • Laser wavelength

      532 nm

    • Spectral coverage

      150-4350 cm-1 (532 nm)

    • Spectral resolution

      5 cm-1 (532 nm)

    • Channels

      Up to four channels

    • Temperature

      Operating:
      15 to 30 °C
      Storage:
      -15 to 50 °C

    • Relative humidity

      20-80% RH, non-condensing

    • Input voltage

      100-240 V, 50-60 Hz, ±10%

    • Power consumption (W)

      400 (max)
      250 (typical start-up)
      120 (typical running)

    • Warm up time (minutes)

      120

    • Unit dimensions (width x height x depth in mm)

      Benchtop model: 279 x 483 x 592
      Cart model: 685 x 1022 (to tabletop) x 753

    • Weight (kg)

      Base model: 32
      Cart model: 93

    • Sampling probe compatibility

      Raman Rxn-30

    • Automation interface

      OPC
      Modbus
      HTTPS
      (contact us for other options)

    • Installation options

      Benchtop (standard) or wheeled cart

    • Hazardous area certifications

      ATEX, CSA, IECEx