起订量:
Level E+H超声波液位计恩德斯豪斯Level
传感器 | Prosonic T(两线制) 一体化 FMU 30 | Prosonic S(四线制) FMU 90/FMU 95(*多带10个探头), 带探头FDU 90...96 | ||||||
FDU 90 | FDU 91 | FDU 91F | FDU 92 | FDU 93 | FDU 95 | FDU 96 | ||
温度°C | -20...+60 | -40...+80 | -40...+80 | -40...+105 | -40...+95 | - 40...+95 | -40...+80 | -40...+150 |
操作压力(abs)bar | 0.7...3 | 0.7...4 | 0.7...4 | 0.7...4 | 0.7...4 | 0.7...3 | 0.7...1.5 | 0.7...3 |
过程连接 | G/NPT 1-1/2'' , 2'' | G1'', 1''NPT | G1'', 1''NPT | G1'', 1''NPT Triclamp, 法兰 | G1'', 1''NPT | G1'', 1''NPT | G1'', 1''NPT | G1'', 1''NPT |
测量端 材质 | PP | PVDF | 整体焊接(无密封圈) | UP, AI/PTFE 套封膜片 | UP, PE-foam SS 膜片 | UP, AI/PTFE 套封膜片 | ||
PVDF | SS316L | PVDF | ||||||
密封材质 | EPDM | - | ||||||
盲区m | 0.25 / 0.35 | 0.07 | 0.3 | 0.3 | 0.4 | 0.6 | 0.7/0.9 | 1.6 |
*大量程(液体)m | 5/8 | 3 | 10 | 10 | 20 | 25 | - | - |
*大量程(固体)m | 2 / 3.5 | 1.2 | 5 | 5 | 10 | 15 | 45 | 70 |
外壳防护等级 | IP66/68 | IP68/NEMA 6P | ||||||
输出 | 4...20 mA | 4...20 mA / HART , PROFIBUS DP, 1, 3或6路继电器或*多4路开关量输入 | ||||||
电源 | 2线制 14...35 VDC | 90...253 VAC 10.5...32 VDC | ||||||
继电器 | FTU 230/231 | 标准一路继电器, 可选3或6路继电器(可自由编程) | ||||||
其它功能 | - | 明渠流量测量, 预置明渠流量曲线, 带积算仪, 泵控制, 耙控制 |
传感器 | Prosonic M(4线制或2线制) 一体化 | |||
FMU40/41 | FMU 42 | FMU 43 | FMU 44 | |
温度°C | -40...+80 | |||
操作压力 (abs)bar | 0.7...3 | 0.7...2.5 | ||
过程连接 | 螺纹 | 法兰, 安装架 | ||
测量端 材质 | PVDF | PVDF | UP / 1.4571 | UP / 1.4571 |
密封 材质 | EPDM | Viton EPDM | EPDM | Viton EPDM |
盲区 m | 0.25 | 0.4 | 0.6 | 0.6 |
*大量程 (液体) m | 5/8 | 10 | 15 | 15 |
*大量程 (固体) m | 2/3.5 | 5 | 7 | 7 |
外壳防护等级 | IP 68/NEMA6P | |||
输出 | 4...20mA / HART , PROFIBUS PA, FF基金会现场总线 | |||
电源 | 2线制: 14...36 VDC 4线制: 90...253 VAC/10.5...36 VDC | |||
其它功能 | 菜单引导式, 包络线 在线分析, 操作工具 ToF Tool, 明渠流量输出 |