Sensor Type | XZR-500-ST + XCU-500-WM |
Measurement Principle | Zirconium Oxide Sensor with Metallic Sealed Reference and K Type T/C |
Performance |
Gas | Wet, dirty and hot |
Measurement Range | 0.01-25% Oxygen |
Accuracy | <2% relative with atmospheric pressure compensation |
Response Time | < 15s for 90% of response step from calibration gas inlet |
Repeatability | ±0.1% |
Fidelity | 1% per month |
Linearity | Better than ±1% |
Sample Flow Rate | Flue Gas at 0.5 m/s minimum rate |
Maximum Sample Pressure | Depending on application |
Maximum Sample Temperature | Up to 1300°C on request |
Outputs |
Output Signal | One off 0/4 to 20 mA linear with Galvanic Insulation Output; 2nd optional output |
Output Load | Over 1000 Ω |
Self-diagnostics | Included in Readout |
Output Ranges | Freely configurated between 0.01-25% |
Alarms | 2 Alarms; User adjustable (10 W) 1 Fault Alarm |
Display Resolution | 0.1% in standard; (0.01% on request) |
Power Supply | 115 or 220 VAC, 50/60 Hz |
Power Consumption | 110 VA |
Ambient Temperature Range | 0°C to 55°C |
Sensor Temperature | 600 to 800°C, optimised at 700°C |
Operating Humidity | 5-90% RH without condensation |
Physical |
Dimensions | 300 x 300 x 200 mm (Control) 290 x 135 x 650 mm (Sensor) |
Weight | 10 to 15kg depending on application |
Wetted Materials | XZR-501- 304 L Stainless steel (700°C) XZR-502- Inconel 600 (1000°C) XZR-503- HR160 (600°C to 1000°C) XZR-504- C2000 (600°C) XZR-505- Halar coating (120°C) XZR-506- Ceramic (1300°C) |
Probe Dimensions | Length 400mm, other on request |
Installation | Stack and wall mounting |
Housing Ingress Protection | Weatherproof |