SARS-CoV-2 N protein ELISA Kit

首页>生物试剂>生物检测试剂>elisa试剂盒

SARS-CoV-2 N protein ELISA Kit

型号
武汉纽斯特生物科技有限公司

免费会员 

生产厂家

该企业相似产品

cAMP ELISA Kit(中文)

在线询价

cAMP EIA Kit

在线询价

SARS-CoV2-N (Full length,原核表达)

在线询价

Human C Reactive Protein (CRP) ELISA Kit

在线询价

Human Transforming Growth Factor Beta 1 (TGFb1) EL...

在线询价

Human Vascular Endothelial Growth Factor A (VEGFA)...

在线询价

Human Interferon Gamma (IFNg) ELISA Kit

在线询价

Human Tumor Necrosis Factor Alpha (TNFa) ELISA Kit

在线询价
cAMP ELISA kit,active抗体,点突变抗体,ELISA试剂盒

武汉纽斯特生物科技有限公司(Wuhan NewEast Biotechonology Co.Ltd),系美国生命科学试剂供应商美国NewEastBio在中国的子公司。公司位于武汉东湖*开发区光谷生物城. 美国NewEastBio专注于生命科学和生物技术领域,主要产品有抗体、小分子、ELISA试剂盒、蛋白等,与世界许多高校和研究所、世界大制药公司、生物科技公司和医院等客户建立了*稳定的合作关系,产品销往各个国家。 武汉纽斯特生物将携手美国NewEastBio,负责美国品牌NewEastBio(www.neweastbio.com)产品在中国市场的推广、销售及售后服务。致力于为客户提供一系列高质量和有较好价格竞争力的美国NewEastBio*产品。并提供专业的技术服务:抗体定制,动物实验等

详细信息

Product: SARS-CoV-2 N protein ELISA Kit

Catalog NO.: NE-xg-E001

Assay Type: Sandwich

Research Area: Infection immunity

Synonyms: SARS-CoV-2 Nucleoprotein; SARS-CoV-2 NP; SARS-CoV-2 Nucleocapsid protein; SARS-Cov-2 NC; SARS-Cov-2 Protein N; Nucleoprotein; NCP;

Detection Range: 125 -8,000 pg/mL

Sensitivity: 38.74 pg/mL

 

Standard Preparation

Reconstitute the Standard with 1.0 mL of Universal Diluent, keep for 10 minutes at room temperature, shake gently(not to foam). The concentration of the

 standard in the stock solution is 8,000 pg/mL. Please prepare 7 tubes containing 0.5 mL Universal Diluent and produce a double dilution series according to the picture shown below. Mix each tube thoroughly

 before the next transfer. Set up 7 points of diluted standard such as 8,000 pg/mL, 4,000 pg/mL, 2,000 pg/mL, 

1,000 pg/mL, 500 pg/mL, 250 pg/mL, 125 pg/mL, and the last EP tube with Universal  Diluent  is the blank as 0 pg/mL.

Typical Data

As the OD values of the standard curve may vary according to the conditions of the objective assay

the experimenter should fit a standard curve for each test. Typical standard curve  provided below is for reference only.

Specificity

This assay has high sensitivity and excellent specificity for detection of SARS-CoV-2 N protein. No significant 

cross-reactivity or interference between SARS-CoV-2 N protein and analogues is observed.

Note: Limited by current skills and knowledge, it is impossible for us to complete the cross-reactivity detection 

between SARS-CoV-2 N protein and all the analogues, therefore, cross reaction may still exist.

 

Recovery

Matrices  listed  below  were  spiked  with  certain  level  of  recombinant SARS-CoV-2 N protein  and  the recovery rates were calculated by comparing the measured value to the expected amount of 

SARS-CoV-2 N protein in samples.

 

Matrix

Recovery range(%)

Average(%)

Serum(n=10)

83-107

95

EDTA plasma(n=10)

82-94

85

Heparin plasma(n=10)

83-99

91


Linearity 

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 

SARS-CoV-2N protein and their serial dilutions. The results were demonstrated by the percentage of calculated 

concentration to the expected.

 

Sample

1:2

1:4

1:8

1:16

Serum(n=10)

83-95%

89-103%

82-99%

79-107%

EDTA plasma(n=10)

92-104%

91-99%

78-98%

80-94%

Heparin plasma(n=10)

86-105%

89-102%

81-97%

85-92%


Precision

Intra-assay Precision (Precision within an assay): Three samples with low, middle and high level SARS-CoV-2 N protein were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): Three samples with low, middle and high level SARS-CoV-2 N protein were tested on 3 different plates, 8 replicates in each plate.

 

CV (%) = SD/meanX100

Intra-Assay: CV< 10%

Inter-Assay: CV< 12%

Stability

The stability of  ELISA kit is determined by the loss rate of activity. The loss rate of  this kit is less than 10% within the expiration date under appropriate storage condition.

 

Main components

37  for 7 days

-20  for 12 months

Average(%)

90

95-100

 

To minimize extra influence on the performance, operation procedures and lab conditions,

especially room temperature, air humidity, incubator temperature should be strictly controlled. It is strongly 

suggested that the same operator performs the whole assay from the beginning to the end.