9610 sc Silica AnalyzerLower Maintenance, Less DowntimeDelivering reliable results that save you critical time and effort.90 days of continuous runtimeOnly two liters of reagent are required for the a......
详细介绍
9610 sc Silica Analyzer—Lower Maintenance, Less Downtime Delivering reliable results that save you critical time and effort. 90 days of continuous runtime Only two liters of reagent are required for the analyzer to perform unattended for up to 90 days; twice as long as the previous analyzer versions. Save time on maintenance The industry’s only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps. Avoid downtime Predictive diagnostic tools, including Hach’s proprietary Prognosys technology, warning LEDs, and high-visibility notification screens let you avoid unplanned downtime. Verify easily with Hach Lab products so you don’t waste time second-guessing Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the analyzer to verify in a lab test.
9610 sc Silica Analyzer—Technical Data Measuring Principle Colorimetric Range 0 to 5000 μg/L as SiO2 Accuracy :0 - 500 μg/L: ±1% or ±1 μg/L of reading, whichever is greater; 500 - 5000 μg/L: ±5%(Tested with Hach reagents) Repeatability ± 0.5 μg/L or ±1% of reading, whichever is greater Lower Limit of Detection (LOD) 0.5 μg/L Response Time Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature Reagent Consumption 2 L of each reagent every 90 days with 15 minute cycle time Operating TemperatureRange 5 - 45 °C (41 - 113 °F) Operating Humidity 5 - 95% relative humidity, noncondensing (indoor use only) Sample Pressure 2 - 87 psi (to Preset Pressure Regulator) Sample Temperature 5 - 50 °C (41 - 122 °F) Sample Flow Rate 55 - 300 mL/min Number of Channels 1, 2, 4, 6; programmable sequence Grab Sample Grab Sample In and Grab Sample Out capability Mounting Wall, panel or table
9610 sc Silica Analyzer—Fitting Type Sample line and sample bypass drain: 6 mm (¼-in.) Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.) Power Requirements (Voltage)100 - 240 V AC, 24 V DC Power Requirements (Hz) 50/60 Hz Output 4 - 20 mA Dimensions (H x W x D) 804 mm x 452 mm x 360 mm Enclosure Rating IP56 / NEMA 4X Certifications CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000),KC (EN 61326-1: 2006) C-tick (EN 61326-1: 2006) cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012) *Subject to change without notice. Silica in the sample reacts with molybdate ions under acidic conditions to form silicomolybdic acid complexes. Addition of citric acid destroys the phosphate complexes. Amino acid reagent is then added to reduce the yellow silicomolybdic acid to an intense blue color, which is proportional to the silica concentration. Test results are measured at 815 nm.
9610 sc Silica Analyzer—Instruments 9610.KTO.S0.A1U 9610 sc Silica Analyzer, open chemistry, 1 channel, 100-240 VAC 9610.KTO.S0.A2U 9610 sc Silica Analyzer, open chemistry, 2 channels, 100-240 V AC 9610.KTO.S0.A4U 9610 sc Silica Analyzer, open chemistry, 4 channels, 100-240 VAC 9610.KTO.S0.A6U 9610 sc Silica Analyzer, open chemistry, 6 channels, 100-240 VAC 9610.KTO.S0.D1U 9610 sc Silica Analyzer, open chemistry, 1 channel, 24 VDC 9610.KTO.S0.D2U 9610 sc Silica Analyzer, open chemistry, 2 channels, 24 VDC 9610.KTO.S0.D4U 9610 sc Silica Analyzer, open chemistry, 4 channels, 24 VDC 9610.KTO.S0.D6U 9610 sc Silica Analyzer, open chemistry, 6 channels, 24 VDC Each Analyzer comes with a Hach reagents set for start up and up to 90 days of use
9610 sc Silica Analyzer—Reagents sc Silica Reagent Set sc Reagent 1 Silica, 2 L sc Reagent 2 Silica, 2 L sc Standard 1 Silica, 2 L sc Reagent 3 Silica, Package sc Reagent 4 Silica, 2 L