HIT Series Test Chamber

HIT Series Test Chamber

参考价: 面议

具体成交价以合同协议为准
2024-11-10 11:20:27
9
产品属性
关闭
北京凯维丰科技发展有限责任公司

北京凯维丰科技发展有限责任公司

免费会员
收藏

组合推荐相似产品

产品简介

MeetconditionsGB10586-89hotandhumidtestchambertechnicalconditionGB2423

详细介绍

Meet conditions

GB10586-89 hot and humid test chamber technical condition

GB2423.2 high temperature test and test B

GB10592-89 high and low temperature test kit technical condition

GB2423.3 constant damp heat test box

Uses and features

This test chamber is an environmental test equipment, which is suitable for the performance test of various products and materials under high and low temperature and humid and hot conditions. The ideal temperature and humidity control mode, the balanced temperature and humidity control mode, can automatically obtain a high precision temperature and humidity test environment. With over-temperature alarm, over-pressure protection, water shortage protection, program setting, computer control.

Configuration and use standard of main parts and components:


1. Chamber:

1) Shell: Electrostatic powder spraying on the surface of cold rolled iron plate;

2) Inner gallbladder: mirror stainless steel plate

3) Insulation layer: super glass wool;

4) Seals: silicone rubber;

5) Circulation system: centrifugal fan, forced air convection;

6) Observation window: hollow heated glass, built-in lighting, can observe the test state of the test object;

7) Test holes: 25 mm cable holes (test holes can be enlarged or reduced as required);

8) Test bracket: stainless steel mesh plate (second layer).

2. Refrigeration components:

1) Refrigeration compressor: original French Taikang refrigeration compressor;

2) Filter: Import

3) Solenoid Valve: Import

4) Condenser: air-cooled condenser;

5) Evaporator: diaphragm air heat exchanger;

6) Refrigerant: R404;

7) Cooling mode: air cooling.

3. Main accessories:

1) Temperature Control Instruments: Customized High and Low Temperature and Humidity Instruments

2) Contactors: People's Electrical Appliances;

3) Solid State Relay: Omron, Japan;

4) Heater: high quality nickel-chromium alloy wire;

5) Temperature sensor: Pt100;

7) Humidity sensor: 0~5V humidity transmitter

8) Humidifier: stainless steel steam humidifier;

9) Water adding mode: automatic water adding.

10) Lighting: Lighting lamp is installed in the observation window of the box door to observe the experiment situation in the chamber more clearly.

4. Safety measures:

1) Safe and reliable grounding protection device;

2) Overtemperature in the studio;

3) Overpressure and overload protection of refrigerators;

4) Water shortage protection.  

5. Provide technical data:

1) Technical data: product qualification certificate, instructions for use, instructions for use of purchased parts, etc.

2) Equipped with two shelves for studio.

6. Conditions of use:

1) Environmental temperature: 5 ~30 C;

2) Environmental humidity: not more than 85% R.H;

3) Air quality: no high concentration of dust and corrosive gas;


Model

Inner Sizemm

Outer Sizemm

HIT-100

400×400×500

950×880×1470

HIT-150

500×500×600

985×932×1562

HIT-250

520×630×780

1060×997×1740

HIT-500

700×800×900

1182×1250×1880

HIT-800

800×1000×1000

1360×1500×1950

HIT-010

1000×1000×1000

1485×1520×2020